Calibration, verification, and adjustment are three distinct activities in instrument management — calibration determines errors, verification confirms whether those errors are acceptable, and adjustment corrects the instrument to reduce errors — and they are often confused, leading to compliance gaps in quality management systems. ISO/IEC 17025:2017 defines these terms clearly, and understanding the distinctions is essential for quality engineers, laboratory managers, and calibration programme managers in Singapore who must satisfy auditors, regulators, and customers that their measurement equipment is properly managed.
This guide explains each term, illustrates the differences with practical examples, and clarifies how they apply in common Singapore industrial and regulatory contexts.
Defining Calibration
The international vocabulary of metrology (VIM, JCGM 200:2012) defines calibration as:
"Operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication."
In plain terms: calibration is the measurement comparison that determines how much an instrument's readings differ from the true value (represented by a traceable reference standard), and by how much. The output of calibration is information — specifically, the errors and measurement uncertainties at each calibration point. Calibration does not, by definition, include any action to correct those errors. It is purely a measurement and reporting activity.
This is why a calibration certificate (from an accredited laboratory like Unitest Instruments under SAC-SINGLAS LA-2023-0845-C) reports measurement results and uncertainty — it tells you what the instrument actually does, not what you want it to do. What you do with that information — accept the instrument, adjust it, or retire it — is a separate decision.
Defining Verification
The VIM defines verification as:
"Provision of objective evidence that a given item fulfils specified requirements."
In the context of measuring instruments, verification is the process of checking whether an instrument meets a specified requirement — typically a specification, a tolerance, or a performance standard. Verification uses the results of calibration (or a simplified check) and compares them against the acceptance criterion to reach a pass/fail conclusion.
A verification can be performed without a full calibration (for example, checking a thermometer against a known ice-water bath to confirm it reads 0°C ± 1°C), or it can follow a calibration (comparing the calibration results against the instrument's specification to determine whether it passes). The difference from calibration is that verification answers the binary question "does this instrument meet the requirement?" whereas calibration answers the quantitative question "what are the errors in this instrument's readings?"
Verification in legal metrology: In Singapore, the Weights and Measures Act and its subsidiary legislation require verification of certain weighing and measuring instruments used in trade (retail scales, petrol pump meters, etc.) by authorised verifiers under the authority of A*STAR's Standards, Productivity and Innovation Board (previously SPRING). This legal metrology verification is distinct from quality-system calibration — it is a statutory activity with specific legal consequences for non-compliance.
Defining Adjustment
The VIM defines adjustment as:
"Set of operations carried out on a measuring system so that it provides prescribed indications corresponding to given values of a quantity to be measured."
Adjustment is the physical or software action taken to reduce or eliminate errors identified by calibration. It is the corrective action, not the measurement. Adjustment might involve turning a zero-adjust screw on a mechanical pressure gauge, resetting the span on a temperature transmitter via HART communication, trimming the offset of a digital thermometer, or reconfiguring the internal calibration of a power analyser.
Critically, adjustment must be followed by calibration — you cannot know whether the adjustment was successful or by how much it improved the instrument's performance without performing a calibration after the adjustment. A calibration certificate that reports only pre-adjustment results does not confirm the instrument's state after adjustment.
How Calibration, Verification, and Adjustment Relate
The typical sequence when an instrument is submitted to a calibration laboratory is:
- Initial calibration (as-found): The instrument is calibrated in its received condition, with no adjustments. The as-found results show how the instrument was performing when it arrived — this is valuable data for your quality system (did the instrument drift significantly since its last calibration?).
- Verification: The as-found calibration results are compared against the instrument's specification (or your required tolerance). If within specification, the instrument passes and no adjustment is needed.
- Adjustment (if required): If the as-found results show the instrument is out of its specification, adjustments are made to correct the errors.
- Post-adjustment calibration (as-left): After adjustment, the instrument is calibrated again to confirm that the adjustment was successful and the instrument now meets its specification. The as-left results are what is reported on the final calibration certificate.
A calibration certificate that shows both as-found and as-left results provides the most complete picture of the instrument's history and the laboratory's work. When reading a certificate, check whether the results shown are as-found (pre-adjustment), as-left (post-adjustment), or both. For more on reading certificates, see our article on how to read a calibration certificate.
Common Confusion Between the Terms in Practice
Several common situations arise where these terms are confused or used incorrectly, leading to quality system gaps.
Using "Verified" When "Calibrated" Is Required
ISO 9001:2015 clause 7.1.5 states that monitoring and measuring resources must be "calibrated or verified" at specified intervals against measurement standards traceable to international or national measurement standards. Some quality managers interpret "verified" as meaning they can perform a simple functional check (e.g. measuring a known reference value with the instrument) in place of a full traceable calibration. This is only acceptable if the verification is performed with a reference standard that is itself traceable, the results are recorded, and the acceptance criterion is defined and documented. A check against an uncalibrated reference, or a visual inspection without measurement, does not satisfy the traceability requirement of ISO 9001 clause 7.1.5 in a metrologically rigorous sense.
Assuming Calibration Includes Adjustment
Some users submit an instrument for calibration expecting to receive it back in its best possible condition. If the laboratory performs calibration only (without adjustment), the instrument is returned with a certificate showing its errors, but those errors have not been corrected. Users should specify to the calibration laboratory whether they want calibration only (reporting of errors), or calibration with adjustment (reporting of errors plus correction). Unitest Instruments discusses service options with customers at the point of enquiry to ensure the right scope is agreed before calibration begins. Contact our team for advice on service options.
Treating a Calibration Label as Evidence of Calibration
A calibration sticker on an instrument shows that it was calibrated, but the sticker alone is not the calibration certificate. The certificate is the technical document that records the actual measurement results, reference standards, and uncertainty. Quality system auditors and regulatory inspectors will ask to see the certificate, not the sticker. See our guide on how to read and understand a calibration certificate for the full requirements.
Which Term Applies in Singapore's Key Regulatory Frameworks?
| Regulatory Framework | Relevant Term | Specific Requirement |
|---|---|---|
| ISO 9001:2015 (clause 7.1.5) | Calibration or verification | Traceable calibration or verification at specified intervals; records retained |
| HSA GMP (pharmaceutical) | Calibration | Critical instruments calibrated against traceable standards; records maintained for GMP inspection |
| Weights and Measures Act (Singapore) | Verification (legal metrology) | Statutory verification of trade measuring instruments by authorised verifiers |
| MOM WSH Act | Maintenance (includes calibration) | Safety-critical measuring instruments maintained in safe and reliable condition |
| IEC 61511 (safety instrumented systems) | Proof testing (includes verification) | Periodic functional tests of safety instruments at SIL-appropriate intervals |
| ISO 14001 (environmental) | Calibration or verification | Monitoring equipment calibrated or verified; records maintained |
In-House Calibration and Verification vs External Accredited Calibration
Some organisations perform in-house calibration or verification using their own reference standards, rather than sending every instrument to an external accredited laboratory. This is acceptable under ISO 9001 and many other frameworks, provided:
- The in-house reference standards are themselves calibrated by an accredited laboratory (such as Unitest Instruments under SAC-SINGLAS accreditation LA-2023-0845-C)
- The in-house calibration procedure is documented and validated
- The personnel performing in-house calibration are competent for the task
- Records of in-house calibrations are maintained
- The uncertainty of the in-house calibration is appropriate for the measurement requirement
For many organisations, a practical approach is to calibrate key reference standards externally at an accredited laboratory, and use those reference standards for in-house verification of working instruments in between external calibration events. This combines the traceability assurance of external accredited calibration with the convenience and cost efficiency of in-house verification. Unitest Instruments can supply and calibrate reference standards for use in in-house calibration programmes — visit our products page or calibration services page for more information.
Summary: Three Terms, Three Activities
- Calibration: The measurement comparison that determines what errors are present in an instrument. The output is a certificate reporting results and uncertainty. Does not correct errors.
- Verification: The assessment of whether calibration results (or a simpler check) meet a specified acceptance criterion. Answers "pass or fail?" Does not correct errors.
- Adjustment: The physical or software correction applied to reduce errors identified by calibration. Must be followed by calibration to confirm effectiveness. Changes the instrument's output.
Understanding these distinctions helps you specify the right service from your calibration provider, interpret certificates correctly, and structure your quality management system's instrument management procedures in a way that satisfies auditors and regulators. If you have questions about which service is right for your instruments, contact the Unitest Instruments team — we respond to enquiries within 2 business hours.
